부메뉴
자료실
[Application]Tranylcypromine
Tranylcypromine
Chromatographic Conditions:
Column: CHIRALPAK AY-RH (150 x 4.6) mm; 5 micron
Mobile phase: Water: ACN: Ammonia (01:99:0.2 v/v/v)
Flow Rate: 0.5 mL/min; Column Temp: 15ºC; Inj. Volume: 5 µL
Conc.: 100 ng/mL;
Diluent: Water: Mobile Phase (10:90; v/v) used for 1.0 mg/mL stock solution preparation and further dilutions mobile phase was used as diluent.
Chromatographic Characteristics:
tR1: 5.9
tR2: 6.4
Rs: 2.2
Mass Spectroscopy: ESI Positive, SIM, M+H: 134 Da
SIM (TIC Spectrum)
댓글 0개